Fisher DLC3100 Digital Level Transmitter - Arrival Photos & Technical Data
Baoman Industrial Technology (Chengdu) Co., Ltd. has recently received a shipment of Fisher DLC3100 digital level transmitters, used for displacer-based level / interface / density measurement, with 4-20 mA + HART output and compatibility with 249 series displacer sensors. Below are arrival photos and technical data for your reference.
I. Arrival Photos

Arrival photo (taken by Baoman Industrial, watermarked, serial information redacted)

Arrival and unboxing detail (taken by Baoman Industrial)

Arrival and unboxing detail (taken by Baoman Industrial)

Arrival and unboxing detail (taken by Baoman Industrial)
II. Brand & Model
| Item | Specification |
|---|---|
| Brand | Fisher (an Emerson brand) |
| Model | DLC3100 digital level transmitter (also DLC3100 SIS safety-instrumented version) |
| Measuring principle | Displacer buoyancy → torque-tube rotation (up to 4.4°) → Hall-effect magnet system sensing and digitization |
| Measured variables | Level, liquid-liquid interface, density |
| Output / communication | 4-20 mA DC (forward / reverse action selectable) + HART (1200 baud FSK), NAMUR NE43 alarm currents |
| Compatible sensors | Various 249 external displacers; also fits third-party displacers from Masoneilan, Yamatake, Foxboro-Eckardt, Honeywell, etc. |
III. Technical Data
The DLC3100 is a digital level transmitter based on torque-tube and Hall-effect sensing, with optional temperature compensation (100 Ω platinum RTD). The transmitter body does not contact the process fluid; wetted parts are determined by the 249 sensor fitted. DLC3100 SIS is the safety-instrumented version with SIS classification and trip alarm current enabled by default.
| Item | Specification |
|---|---|
| Measurement range | Depends on fitted displacer length (standard 14–120 in); 4-20 mA can represent a level span up to 2000 m |
| Input signal | Torque-tube shaft rotation; optional 100 Ω platinum RTD (2/3-wire) for temperature compensation |
| Output signal | 4-20 mA DC (forward / reverse action selectable) + HART; NAMUR NE43 alarm currents |
| Power supply | 12-30 V DC, 25 mA; reverse-polarity protection; HART requires a minimum steady 17.75 V |
| Reference accuracy | Independent linearity ±0.25% (NPS3, 14 in displacer) to ±0.8% (other 249) / ±0.5% (249W); repeatability ±0.1%–±0.5% FS; hysteresis <0.2% of output; dead band <0.05% of input |
| Process temperature | Determined by sensor material (cast iron -29 to 232 °C; steel / stainless / nickel-copper -198 to 427 °C); torque tube N05500 not recommended above 232 °C |
| Ambient temperature | -40 to 80 °C; LCD operating limit -20 to +70 °C |
| Enclosure rating | IP66, Type 4X |
| Electrical connections | Two 1/2-14 NPT internal conduit connections (at bottom of terminal box) |
| Weight | ≤ 3.45 kg / 7.57 lb |
| Certifications | CSA (exida) SIL2; ATEX / IECEx, NEPSI, CML, CUTR, etc.; DLC3100 SIS additionally has SIS classification |
Source: Fisher DLC3100 digital level transmitter Instruction Manual D104213X0CN; ordering guide D104216X0CN; certification supplements CSA D104232X0CN, ATEX/IECEx D104233X0CN. Parameters are subject to the latest official documentation.
IV. Related Models
| Item | Specification |
|---|---|
| DLC3100 SIS | Safety-instrumented system (SIS) version with SIS classification for safety-related applications |
| DLC3020f | Same-platform digital level controller with FOUNDATION fieldbus, combining transmitter / controller / switch, accuracy ±0.15% |
| 249 series | Matching displacer sensors (249B/C/K/L, etc.) for level / interface / density sensing |
| L2 / 2500 series | Pneumatic / electric level controllers and pneumatic transmitters for different energy and emission requirements |
V. Selection Guide & Common Models
Selection guide:Selection of the DLC3100 starts with the measured variable (level / interface / density) and the fitted 249 sensor model, displacer length and material, then the torque-tube material and wall thickness, analog output action (forward / reverse), communication version (HART 5 / HART 7) and hazardous-area certification. The transmitter body does not contact the process fluid — wetted materials are determined by the 249 sensor; RTD temperature compensation is optional. Please refer to official guide D104216X0CN and confirm details at inquiry.
Common models:DLC3100 standard (4-20 mA + HART), DLC3100 SIS (SIS safety version), with matching 249B / 249C / 249K / 249L displacer sensor combinations.
These are some of the configurations we supply regularly. If you are working through a selection, or would like to know about availability of these models, feel free to get in touch — we can look at the right combination together based on your operating conditions.
Disclaimer: The information above is not guaranteed to remain valid and may be modified without separate notice. Please confirm the latest information at the time of inquiry.